home *** CD-ROM | disk | FTP | other *** search
/ Shareware Overload Trio 2 / Shareware Overload Trio Volume 2 (Chestnut CD-ROM).ISO / dir26 / med9410o.zip / M94A2724.TXT < prev    next >
Text File  |  1994-10-25  |  2KB  |  33 lines

  1.        Document 2724
  2.  DOCN  M94A2724
  3.  TI    Sample Addition Monitor gives improved safety of blood supply with
  4.        Wellcozyme HIV 1+2 assay.
  5.  DT    9412
  6.  AU    Povey J; Beckford U; Rayment F; Evans CJ; Murex Diagnostics Ltd.,
  7.        Dartford, Kent, U.K.
  8.  SO    Int Conf AIDS. 1994 Aug 7-12;10(1):239 (abstract no. PB0386). Unique
  9.        Identifier : AIDSLINE ICA10/94369851
  10.  AB    OBJECTIVE: Errors in addition of sample, manually or with automatic
  11.        dispensers, affect the sensitivity of assays. A sample diluent was
  12.        therefore developed, which would confirm that sample had been added to
  13.        the well, thereby confirming that the protocol was followed. RESULTS:
  14.        The diluent is green/brown and changes to blue when serum or plasma is
  15.        added to it. This change is easily seen and can be recorded
  16.        spectrophotometrically (absorbance increases from 0.05 to 0.35). Assay
  17.        sensitivity has been improved with the following bleeds detected of BBI
  18.        HIV-1 seroconversion panels; BBI R3, L2, Q5, I2, S2 and of NABI panels;
  19.        151-2C, 171-2 C and 121-2C. The sensitivity of HIV-2 detection has also
  20.        been increased. CONCLUSION: The inclusion of a Sample Addition Monitor
  21.        facilitates the confirmation that sample has been added to the well.
  22.        Mistakes can be easily identified and rectified thereby improving the
  23.        safety of blood supply.
  24.  DE    AIDS Serodiagnosis/*INSTRUMENTATION  Blood Banks  *Blood Donors  Human
  25.        HIV Seropositivity/*DIAGNOSIS/IMMUNOLOGY  HIV-1/*IMMUNOLOGY
  26.        HIV-2/*IMMUNOLOGY  Immunoenzyme Techniques/*INSTRUMENTATION  Predictive
  27.        Value of Tests  Quality Control  Specimen Handling/INSTRUMENTATION
  28.        Spectrophotometry/INSTRUMENTATION  MEETING ABSTRACT
  29.  
  30.        SOURCE: National Library of Medicine.  NOTICE: This material may be